| ENGINEERING SPECIFICATIONS: StarTOC AUTOMATIC BENCHTOP TOC | ||||||||
| Model/Name | StarTOC Automatic Benchtop Analyzer | Notes | ||||||
| Analysis Methods | All methods in single Analyzer: UV/Heated Persulfate, High Temperature Combustion, Ozone Promoted/Hydroxyl Radical, Heated Persulfate, Ultra-Pure (UV - NDIR) | Operator selectable - No Hardware Change | ||||||
| Analytes Measured | Total Organic Carbon; Total Carbon; Total Inorganic Carbon; Non-Purgeable Organic Carbon; TOC-True (including purgeable Organics) | |||||||
| Detector Type (CO2) | NDIR (solid state; no moving parts; computer-controlled; non-reflective sample cell-impervious to corrosion and guaranteed for 5 years) | Footnote 1 - NDIR vs Conductivity | ||||||
| Control/Data Handling | Microsoft Windows PC | |||||||
| Sample Introduction | Digital Injection/Auto-Sampler | |||||||
| Measurement Range (mg/L) | 0.100 - 50,000 | Dependent on Chosen Method | ||||||
| No Operator Dilution | ||||||||
| Accuracy/Repeatability (%) | £5% RSD or 2% FS (whichever is greater) | |||||||
| Carrier Gas Flow (mL/min.) | 300 mL/max - Computer Controlled Mass Flow | Bottled gas preferred for low | ||||||
| Controller CO2 & HC - FREE AIR, OR O2 | ppm/ppb measurements | |||||||
| 15 +/- 2 PSI | ||||||||
| Average Analysis Time (minutes) | 3 | |||||||
| Outputs | RS-232 | |||||||
| Display | Flat Color Screen (external PC) | |||||||
| Data Storage | External PC | |||||||
| UTILITIES Required | ||||||||
| Power: | 100/200 VAC | |||||||
| 50/60 HZ. | ||||||||
| 15 Amp Service | ||||||||
| Carrier Gas | CO2 & HC - free air, or O2 (300 mL/minute-max.); | |||||||
| 15 +/- 2 PSI | ||||||||
| Reagents | Prepackaged Sodium Persulfate | |||||||
| Phosphoric Acid | ||||||||
| Calibration Standards | ||||||||
| D.I. Water | ||||||||
| Sample Drain | gravity/air break | |||||||
| Environment | Operating Temperature: 10° - 50°C | |||||||
| 50° - 122°F | ||||||||
| Construction | ||||||||
| Enclosure | Epoxy Powder Coated cold rolled Steel | |||||||
| Dimensions (HXWXD) | 51 X 51 X 38 (cm) | |||||||
| 25 x 22 x 24 (in) | ||||||||
| Mounting | Benchtop | |||||||
| Weight | 59 Kg | |||||||
| 131 Lbs | ||||||||
| Compliance/Certification | CE, CSA, ASTM, EPA Methods | |||||||
| Conformity | Complying with all International Standards, such as: | |||||||
| DIN-EN 1484, DIN-ENV 12260, DIN 38409-H3, | ||||||||
| ISO 8245, Standard Method 5310B, Standard Method | ||||||||
| 5310C, Standard Method 5310D, USEPA 415, | ||||||||
| USEPA 9060, ASTM D5173, EN 13137 | ||||||||
| General Features | Auto-Calibration, | |||||||
| Auto-Validation, | ||||||||
| Auto-Cleaning, | ||||||||
| Historical Data/Time and Date Stamped | ||||||||
| Options | ||||||||
| Total Nitrogen | ||||||||
| Oxygen Generator (electricity only) | Eliminates Gas Bottles for up | |||||||
| to 5 analyzers. Plant Air Not | ||||||||
| Required. (For indoor/climate | ||||||||
| controlled environments only) | ||||||||
| Printer | ||||||||
| (All performance specifications have been verified in a controlled laboratory environment. Actual field performance may vary with application. Measuring range and detection limits depend on the method, injection volume, vessel purity, chemicals and gases used, and the qualification of the operators. | ||||||||
| Footnote 1 | ||||||||
| Conductivity vs. NDIR Interferences Chart | ||||||||
| Reliability of TOC Results in ULTRAPURE WATER ANALYSIS | ||||||||